Setting the Peak Phase Error Limit


You can set an upper Peak Phase Error Limit for your Measurement results.
If Peak Ph Err Limit is turned On, and your measured Peak Phase Error is below the limit you have specified, the numeric results reported in the metrics portion of the screen will be displayed in green, and the Limit Pass will be indicated by "(P)" after the result. If your measured Peak Phase Error is above your specified limit, the numeric result reported in the metrics portion of the screen will be displayed in red, and the Limit Failure will be indicated by "(F)" after the results.
To Set the Peak Phase Error Limit:

Step
Notes
1
[Setup]

2
[Peak Ph Err Limits]

3
[Limits]

4
[On] or [Off]

5
Set the high limit.
  1. [High Limit]
  2. Enter the number on the numeric keypad.
  3. [deg]