| |
|
|
|
Step
|
Notes
|
1
|
Mode
|
|
2
|
[Power Meter]
|
|
3
|
[Zero]
|
Follow the Zeroing Wizard, which displays instructions on the screen.
|
4
|
[Continue]
|
|
5
|
[Ref CF]
|
|
6
|
Enter the reference cal factor using the numeric keypad.
|
The reference cal factor is printed on the label of the power sensor head.
|
7
|
[%]
|
|
8
|
[Calibrate]
|
|
9
|
Connect the power sensor to the Power Ref 50 MHz port.
|
The test set supports the Agilent 8482A and 8481A power sensors.
|
10
|
[Continue]
|
|
11
|
[Cal Factor]
|
|
12
|
Enter the cal factor using the numeric keypad.
|
A list of cal factors is printed on the label of the power sensor head. Select one that's within the operating frequency of the base station.
For an example of a power sensor calibration table, see below.
Using the example power sensor calibration table, if you were interested in 800 MHz, then:
800 MHz has a Cal Factor percentage of ~98.2
Note: Interpolate the Cal Factor from the frequency range provided on the power sensor. The example sensor calibration table shows that 800 MHz is between 300 to 1000 MHz. Since 300 MHz is listed at 98.5 and 1000 MHz (1 GHz) is at 98.1, 800 MHz would be ~98.2.
|
13
|
[%]
|
|
14
|
Connect the external attenuator.
|
|
15
|
If you have previously stored PM loss from an insertion loss measurement, go to the next step.
Or
Press [PM Loss] and enter the PM loss value.
|
For information on the effects of the PM Loss function on the measurement, refer to "Using the Power Meter Loss field, below.
|
16
|
Connect to the base station.
|
Connect the power sensor to the signal to be measured. The connection is typically after the output of the power amplifier or duplexer. The connection requires, in addition to the power sensor and attenuator, a 2 ft. cable and two Type-N barrel connectors.
|